CVE-2026-17583

Thermo Fisher Applied Biosystems Genetic Analyzers Missing Support for Integrity Check

HIGH
8.4
CVSS Score
Published: August 5, 2026 at 08:23 PM UTC
Updated: August 26, 2026 at 02:50 PM UTC
14 products affected across 1 vendorApplied Biosystems 3500 /3500xl Series Data Collection SoftwareABI PRISM 310 Data Collection SoftwareApplied Biosystems 3730 /3730xl Series Data Collection Software

Using Applied Biosystems 3500/3500xL Series Data Collection Software by Thermo Fisher?

Get instant alerts when new vulnerabilities affect Applied Biosystems 3500/3500xL Series Data Collection Software - before attackers find them.

Real-time alertsTrack your software stackFree plan available
Severity
HIGH8.4
Exploit prediction
EPSS 0.1%
SSVC
No known exploit

Description

The affected Thermo Fisher Applied Biosystems Genetic Analyzers are vulnerable because .fsa/.hid output files can be edited. An attacker could tamper with these files, altering DNA data and resulting in inaccurate DNA test outcomes.

Remediation

No guide yet

No remediation guidance has been written for this CVE yet. What the record does show:

References (3)

PatchThird Party AdvisoryUS Government Resource
Source: cverecord
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-216-01
Third Party Advisory
Source: cverecord
https://www.cve.org/CVERecord?id=CVE-2026-17583

SSVC — Action Prioritization

SSVCNo known exploitNot automatableTotal impact

SSVC signals action urgency (what to patch first), separate from CVSS which measures severity.

CVSS

8.4HIGHicscert
Attack Vector
LOCAL
Complexity
LOW
Privileges
NONE
User Interaction
NONE
Scope
UNCHANGED
Confidentiality
HIGH
Integrity
HIGH
Availability
HIGH
Vector?

Quick Info

StatusANALYZED
CWE
CWE-353

Get alerted

Stop reading CVE feeds looking for your own stack

Track the products you run and VulnTracker tells you when a new CVE affects them — with what is known about fixing it, and nothing about the rest.

Free plan, no card