Unauthenticated Remote Code Execution via .NET Deserialization
Get instant alerts when new vulnerabilities affect IntelliSpace Portal - before attackers find them.
The IntelliSpace portal application utilizes .NET Remoting for its functionality. The vulnerability arises from the exploitation of port 755 through the deserialization vulnerability. After analyzing the configuration files, we observed that the server had set the TypeFilterLevel to Full which is dangerous as it can potentially lead to remote code execution using deserialization. This issue affects IntelliSpace Portal: 12 and prior.
Our own answer: ranges merged, affected separated from platform-only, with confidence and review flags.
No remediation guidance has been written for this CVE yet, and no fixed version or vendor advisory is recorded against it.
SSVC signals action urgency (what to patch first), separate from CVSS which measures severity.
Get alerted
Track the products you run and VulnTracker tells you when a new CVE affects them — with what is known about fixing it, and nothing about the rest.