CVE-2025-14095

Privilege boundary violation in Radiometer Products

MEDIUM
6.8
CVSS Score
Published: December 17, 2025 at 11:45 AM UTC
Updated: October 7, 2026 at 11:10 AM UTC
6 products affected across 1 vendorAqt90-flexAbl800-flexAbl800-basic

Using ABL90 FLEX and ABL90 FLEX PLUS Analyzers by Radiometer Medical Aps?

Get instant alerts when new vulnerabilities affect ABL90 FLEX and ABL90 FLEX PLUS Analyzers - before attackers find them.

Real-time alertsTrack your software stackFree plan available
Severity
MEDIUM6.8
Exploit prediction
EPSS 0.2%
SSVC
No known exploit

Description

A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software.                                                                                                                                                                                                  Other related CVE's are CVE-2025-14096 and CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required configuration for Exposure:  Physical access to the analyzer is needed. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution:                                                                                                                                                                                                                                                                                                                                                                              Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status:                                                                                                                                                                                                                                                                                                                                                                                       Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication.

Affected

Our own answer: ranges merged, affected separated from platform-only, with confidence and review flags.

v < 6.20
Affected range stops below 6.20, so it looks safe — our inference, not a vendor statement
v < 6.20
Affected range stops below 6.20, so it looks safe — our inference, not a vendor statement
v < 1.5.0
Affected range stops below 1.5.0, so it looks safe — our inference, not a vendor statement
Affected range stops below 3.5MR11, so it looks safe — our inference, not a vendor statement
Affected range stops below 3.5MR11, so it looks safe — our inference, not a vendor statement

Remediation

No guide yet

No remediation guidance has been written for this CVE yet, and no fixed version or vendor advisory is recorded against it.

References (1)

SSVC — Action Prioritization

SSVCNo known exploitNot automatablePartial impact

SSVC signals action urgency (what to patch first), separate from CVSS which measures severity.

CVSS

6.8MEDIUMRadiometer
Attack Vector
PHYSICAL
Complexity
LOW
Privileges
NONE
User Interaction
NONE
Scope
UNCHANGED
Confidentiality
HIGH
Integrity
HIGH
Availability
HIGH
Vector?

Quick Info

StatusDEFERRED
CWE
CWE-284CWE-693

Get alerted

Stop reading CVE feeds looking for your own stack

Track the products you run and VulnTracker tells you when a new CVE affects them — with what is known about fixing it, and nothing about the rest.

Free plan, no card