CVE-2025-1204

HIGH
7.7
CVSS Score
Published: February 25, 2025 at 05:00 PM UTC
Updated: June 17, 2026 at 08:38 AM UTC
1 product affected across 1 vendorCMS8000

Using CMS8000 Patient Monitor by Contec Health?

Get instant alerts when new vulnerabilities affect CMS8000 Patient Monitor - before attackers find them.

Real-time alertsTrack your software stackFree plan available
Severity
HIGH7.7
Exploit prediction
EPSS 0.5%
SSVC
PoC

Description

The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.

Affected

Our own answer: ranges merged, affected separated from platform-only, with confidence and review flags.

All versions

Remediation

No guide yet

No remediation guidance has been written for this CVE yet, and no fixed version or vendor advisory is recorded against it.

SSVC — Action Prioritization

SSVCPoCNot automatableTotal impact

SSVC signals action urgency (what to patch first), separate from CVSS which measures severity.

CVSS

7.7HIGHicscert
Vector?

Quick Info

StatusDEFERRED
CWE
CWE-912

Get alerted

Stop reading CVE feeds looking for your own stack

Track the products you run and VulnTracker tells you when a new CVE affects them — with what is known about fixing it, and nothing about the rest.

Free plan, no card