CVE-2025-0626

Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor

HIGH
7.5
CVSS Score
Published: January 30, 2025 at 06:17 PM UTC
Updated: June 17, 2026 at 08:26 AM UTC
1 product affected across 1 vendorCMS8000

Using CMS8000 Patient Monitor by Contec Health?

Get instant alerts when new vulnerabilities affect CMS8000 Patient Monitor - before attackers find them.

Real-time alertsTrack your software stackFree plan available
Severity
HIGH7.5
Exploit prediction
EPSS 1%
SSVC
No known exploit

Description

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

Affected

Our own answer: ranges merged, affected separated from platform-only, with confidence and review flags.

All versions

Remediation

No guide yet

No remediation guidance has been written for this CVE yet, and no fixed version or vendor advisory is recorded against it.

SSVC — Action Prioritization

SSVCNo known exploitNot automatableTotal impact

SSVC signals action urgency (what to patch first), separate from CVSS which measures severity.

CVSS

7.5HIGHicscert
Attack Vector
NETWORK
Complexity
HIGH
Privileges
NONE
User Interaction
REQUIRED
Scope
UNCHANGED
Confidentiality
HIGH
Integrity
HIGH
Availability
HIGH
Vector?

Quick Info

StatusDEFERRED
CWE
CWE-912

Get alerted

Stop reading CVE feeds looking for your own stack

Track the products you run and VulnTracker tells you when a new CVE affects them — with what is known about fixing it, and nothing about the rest.

Free plan, no card