CVE-2023-30561

Lack of Cryptographic Security of IUI Bus

MEDIUM
6.1
CVSS Score
Published: July 13, 2023 at 07:03 PM UTC
Updated: June 17, 2026 at 05:55 AM UTC
3 products affected across 2 vendorsAlaris 8015 Pcu FirmwareAlaris 8015-pcuAlaris 8015-pcu

Using BD Alarisâ„¢ Point-of-Care Unit (PCU) Model 8015 by Becton Dickinson & Co?

Get instant alerts when new vulnerabilities affect BD Alarisâ„¢ Point-of-Care Unit (PCU) Model 8015 - before attackers find them.

Real-time alertsTrack your software stackFree plan available
Severity
MEDIUM6.1
Exploit prediction
EPSS 0.2%
SSVC
No known exploit

Description

The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device while an infusion is running.

Affected

Our own answer: ranges merged, affected separated from platform-only, with confidence and review flags.

v ≤ 12.1.3
bd-alaris-8015-pcu (same device)
All versions

Remediation

No guide yet

No remediation guidance has been written for this CVE yet. What the record does show:

SSVC — Action Prioritization

SSVCNo known exploitNot automatablePartial impact

SSVC signals action urgency (what to patch first), separate from CVSS which measures severity.

CVSS

6.1MEDIUMBD
Attack Vector
PHYSICAL
Complexity
LOW
Privileges
NONE
User Interaction
NONE
Scope
UNCHANGED
Confidentiality
HIGH
Integrity
NONE
Availability
HIGH
Vector?

Quick Info

StatusMODIFIED
CWE
CWE-311

Get alerted

Stop reading CVE feeds looking for your own stack

Track the products you run and VulnTracker tells you when a new CVE affects them — with what is known about fixing it, and nothing about the rest.

Free plan, no card