CVE-2022-38100

Contec Health CMS8000

HIGH
7.5
CVSS Score
Published: September 13, 2022 at 02:54 PM UTC
Updated: June 17, 2026 at 04:56 AM UTC
2 products affected across 1 vendorCMS8000 FirmwareCMS8000

Using CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor by Contec Health?

Get instant alerts when new vulnerabilities affect CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor - before attackers find them.

Real-time alertsTrack your software stackFree plan available
Severity
HIGH7.5
Exploit prediction
EPSS 1.0%
SSVC
No known exploit

Description

The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.

Affected

Our own answer: ranges merged, affected separated from platform-only, with confidence and review flags.

All versions
contechealth-cms8000 (same device)
All versions

Remediation

No guide yet

No remediation guidance has been written for this CVE yet. What the record does show:

References (1)

x_refsource_MISCMitigationThird Party AdvisoryUS Government Resource
https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01

SSVC — Action Prioritization

SSVCNo known exploitAutomatablePartial impact

SSVC signals action urgency (what to patch first), separate from CVSS which measures severity.

CVSS

7.5HIGHicscert
Attack Vector
NETWORK
Complexity
LOW
Privileges
NONE
User Interaction
NONE
Scope
UNCHANGED
Confidentiality
NONE
Integrity
NONE
Availability
HIGH
Vector?

Quick Info

StatusMODIFIED
CWE
CWE-400NVD-CWE-Other

Get alerted

Stop reading CVE feeds looking for your own stack

Track the products you run and VulnTracker tells you when a new CVE affects them — with what is known about fixing it, and nothing about the rest.

Free plan, no card