CVE-2020-27260

MEDIUM
5.3
CVSS Score
Published: January 8, 2021 at 03:17 PM UTC
Updated: June 17, 2026 at 03:09 AM UTC
2 products affected across 1 vendorVital Signs Monitor VC150 FirmwareVital Signs Monitor Vc150

Using BIGCOMPANYSOFT SOFTWARE PRODUCT by N/A?

Get instant alerts when new vulnerabilities affect BIGCOMPANYSOFT SOFTWARE PRODUCT - before attackers find them.

Real-time alertsTrack your software stackFree plan available
Severity
MEDIUM5.3
Exploit prediction
EPSS 0.4%

Description

Innokas Yhtymä Oy Vital Signs Monitor VC150 prior to Version 1.7.15 HL7 v2.x injection vulnerabilities exist in the affected products that allow physically proximate attackers with a connected barcode reader to inject HL7 v2.x segments into specific HL7 v2.x messages via multiple expected parameters.

Affected

Our own answer: ranges merged, affected separated from platform-only, with confidence and review flags.

v < 1.7.15
Affected range stops below 1.7.15, so it looks safe — our inference, not a vendor statement
innokasmedical-vital-signs-monitor-vc150 (same device)

Remediation

No guide yet

No remediation guidance has been written for this CVE yet, and no fixed version or vendor advisory is recorded against it.

References (1)

x_refsource_MISCThird Party AdvisoryUS Government Resource
https://us-cert.cisa.gov/ics/advisories/icsma-21-007-01

CVSS

Attack Vector
PHYSICAL
Complexity
LOW
Privileges
NONE
User Interaction
NONE
Scope
CHANGED
Confidentiality
NONE
Integrity
HIGH
Availability
NONE
Vector?

Quick Info

StatusMODIFIED
CWE
CWE-74

Get alerted

Stop reading CVE feeds looking for your own stack

Track the products you run and VulnTracker tells you when a new CVE affects them — with what is known about fixing it, and nothing about the rest.