CVE-2020-12047

CRITICAL
9.8
CVSS Score
Published: June 29, 2020 at 01:41 PM UTC
Updated: June 17, 2026 at 02:51 AM UTC
3 products affected across 1 vendorSigma Spectrum Infusion System FirmwareSigma Spectrum Infusion SystemWireless Battery Module

Using Baxter Sigma Spectrum Infusion Pumps by N/A?

Get instant alerts when new vulnerabilities affect Baxter Sigma Spectrum Infusion Pumps - before attackers find them.

Real-time alertsTrack your software stackFree plan available
Severity
CRITICAL9.8
Exploit prediction
EPSS 2%

Description

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with hard-coded credentials.

Affected

Remediation

No guide yet

No remediation guidance has been written for this CVE yet, and no fixed version or vendor advisory is recorded against it.

References (1)

x_refsource_MISCThird Party AdvisoryUS Government Resource
https://www.us-cert.gov/ics/advisories/icsma-20-170-04