CVE-2020-10627

HIGH
8.1
CVSS Score
Published: December 1, 2021 at 03:38 PM UTC
Updated: June 17, 2026 at 02:48 AM UTC
2 products affected across 1 vendorOmnipod Insulin Management SystemOmnipod Insulin Management System Firmware

Using Omnipod Insulin Management System by Insulet?

Get instant alerts when new vulnerabilities affect Omnipod Insulin Management System - before attackers find them.

Real-time alertsTrack your software stackFree plan available
Severity
HIGH8.1
Exploit prediction
EPSS 0.5%

Description

Insulet Omnipod Insulin Management System insulin pump product ID 19191 and 40160 is designed to communicate using a wireless RF with an Insulet manufactured Personal Diabetes Manager device. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with access to one of the affected insulin pump models may be able to modify and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

Affected

Our own answer: ranges merged, affected separated from platform-only, with confidence and review flags.

insulet-omnipod-insulin-management-system (same device)

Remediation

No guide yet

No remediation guidance has been written for this CVE yet. What the record does show:

References (2)

x_refsource_MISCThird Party AdvisoryUS Government Resource
https://us-cert.cisa.gov/ics/advisories/icsma-20-079-01

CVSS

Attack Vector
ADJACENT_NETWORK
Complexity
LOW
Privileges
NONE
User Interaction
NONE
Scope
UNCHANGED
Confidentiality
HIGH
Integrity
HIGH
Availability
NONE
Vector?

Quick Info

StatusMODIFIED
CWE
CWE-284NVD-CWE-Other

Get alerted