Medtronic Conexus Radio Frequency Telemetry Protocol Cleartext Transmission of Sensitive Information
Get instant alerts when new vulnerabilities affect Conexus Radio Frequency Telemetry Protocol - before attackers find them.
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.
Our own answer: ranges merged, affected separated from platform-only, with confidence and review flags.
No remediation guidance has been written for this CVE yet, and no fixed version or vendor advisory is recorded against it.
Get alerted