CVE-2019-10966

MEDIUM
5.3
CVSS Score
Published: July 10, 2019 at 05:52 PM UTC
Updated: June 17, 2026 at 02:12 AM UTC
8 products affected across 1 vendorAestiva 7900 FirmwareAespire 7100 FirmwareAestiva 7100 Firmware

Using Aestiva and Aespire by GE?

Get instant alerts when new vulnerabilities affect Aestiva and Aespire - before attackers find them.

Real-time alertsTrack your software stackFree plan available
Severity
MEDIUM5.3
Exploit prediction
EPSS 1%

Description

In GE Aestiva and Aespire versions 7100 and 7900, a vulnerability exists where serial devices are connected via an added unsecured terminal server to a TCP/IP network configuration, which could allow an attacker to remotely modify device configuration and silence alarms.

Affected

Our own answer: ranges merged, affected separated from platform-only, with confidence and review flags.

All versions
ge-aespire-7100 (same device)
All versions
ge-aespire-7900 (same device)
All versions
ge-aestiva-7100 (same device)
All versions
ge-aestiva-7900 (same device)

Remediation

No guide yet

No remediation guidance has been written for this CVE yet. What the record does show:

References (2)

VDB Entryx_refsource_BIDThird Party Advisory
http://www.securityfocus.com/bid/109102
x_refsource_MISCMitigationThird Party AdvisoryUS Government Resource
https://www.us-cert.gov/ics/advisories/icsma-19-190-01

CVSS

Attack Vector
NETWORK
Complexity
LOW
Privileges
NONE
User Interaction
NONE
Scope
UNCHANGED
Confidentiality
NONE
Integrity
LOW
Availability
NONE
Vector?

Quick Info

StatusMODIFIED
CWE
CWE-287

Get alerted

Stop reading CVE feeds looking for your own stack

Track the products you run and VulnTracker tells you when a new CVE affects them — with what is known about fixing it, and nothing about the rest.

Free plan, no card