Medtronic 2090 Carelink Programmer Improper Restriction of Communication Channel to Intended Endpoints
Get instant alerts when new vulnerabilities affect 2090 CareLink Programmer - before attackers find them.
Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.
Our own answer: ranges merged, affected separated from platform-only, with confidence and review flags.
No remediation guidance has been written for this CVE yet, and no fixed version or vendor advisory is recorded against it.