CVE-2017-9658

N/A
Published: April 30, 2018 at 03:00 PM UTC
Updated: June 17, 2026 at 01:28 AM UTC
2 products affected across 1 vendorIntellivue MX40 FirmwareIntellivue Mx40

Using IntelliVue MX40 Patient Worn Monitor by Philips?

Get instant alerts when new vulnerabilities affect IntelliVue MX40 Patient Worn Monitor - before attackers find them.

Real-time alertsTrack your software stackFree plan available
Severity
N/A
Exploit prediction
EPSS 0.8%

Description

Certain 802.11 network management messages have been determined to invoke wireless access point blacklisting security defenses when not required, which can necessitate intervention by hospital staff to reset the device and reestablish a network connection to the Wi-Fi access point. During this state, the Philips IntelliVue MX40 Version B.06.18 can either connect to an alternative access point within signal range for association to a central monitoring station, or it can remain in local monitoring mode until the device is reset by hospital staff. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

Affected

Our own answer: ranges merged, affected separated from platform-only, with confidence and review flags.

Affected range stops below b.06.18, so it looks safe — our inference, not a vendor statement
philips-intellivue-mx40 (same device)
All versions

Remediation

No guide yet

No remediation guidance has been written for this CVE yet. What the record does show:

References (3)

VDB Entryx_refsource_BIDThird Party Advisory
http://www.securityfocus.com/bid/100813
x_refsource_MISCThird Party AdvisoryUS Government Resource
https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01

CVSS

Confidentiality
NONE
Integrity
NONE
Availability
COMPLETE
Vector?

Quick Info

StatusMODIFIED
CWE
CWE-755

Get alerted

Stop reading CVE feeds looking for your own stack

Track the products you run and VulnTracker tells you when a new CVE affects them — with what is known about fixing it, and nothing about the rest.

Free plan, no card