CVE-2017-9657

N/A
Published: April 30, 2018 at 03:00 PM UTC
Updated: June 17, 2026 at 01:28 AM UTC
2 products affected across 1 vendorIntellivue MX40 FirmwareIntellivue Mx40

Using IntelliVue MX40 Patient Worn Monitor by Philips?

Get instant alerts when new vulnerabilities affect IntelliVue MX40 Patient Worn Monitor - before attackers find them.

Real-time alertsTrack your software stackFree plan available
Severity
N/A
Exploit prediction
EPSS 0.8%

Description

Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

Affected

Our own answer: ranges merged, affected separated from platform-only, with confidence and review flags.

Affected range stops below b.06.18, so it looks safe — our inference, not a vendor statement
philips-intellivue-mx40 (same device)
All versions

Remediation

No guide yet

No remediation guidance has been written for this CVE yet. What the record does show:

References (3)

VDB Entryx_refsource_BIDThird Party Advisory
http://www.securityfocus.com/bid/100813
x_refsource_MISCThird Party AdvisoryUS Government Resource
https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01

CVSS

Confidentiality
NONE
Integrity
NONE
Availability
PARTIAL
Vector?

Quick Info

StatusMODIFIED
CWE
CWE-460CWE-755