CVE-2017-12714

N/A
Published: April 25, 2018 at 01:00 PM UTC
Updated: June 17, 2026 at 01:03 AM UTC
20 products affected across 2 vendorsAccent St FirmwareAccent Mri FirmwareAccent Firmware

Using Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI. by Abbott Laboratories?

Get instant alerts when new vulnerabilities affect Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI. - before attackers find them.

Real-time alertsTrack your software stackFree plan available
Severity
N/A
Exploit prediction
EPSS 0.6%

Description

Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

Affected

Our own answer: ranges merged, affected separated from platform-only, with confidence and review flags.

v < f0b.0e.7e
Affected range stops below f0b.0e.7e, so it looks safe — our inference, not a vendor statement
abbott-accent (same device)
v < f10.08.6c
Affected range stops below f10.08.6c, so it looks safe — our inference, not a vendor statement
abbott-accent-mri (same device)
v < f10.08.6c
Affected range stops below f10.08.6c, so it looks safe — our inference, not a vendor statement
abbott-accent-st (same device)
v < f14.07.80
Affected range stops below f14.07.80, so it looks safe — our inference, not a vendor statement
abbott-allure (same device)
v < f0b.0e.7e
Affected range stops below f0b.0e.7e, so it looks safe — our inference, not a vendor statement
abbott-anthem (same device)
v < f14.07.80
Affected range stops below f14.07.80, so it looks safe — our inference, not a vendor statement
abbott-assurity (same device)
v < f17.01.49
Affected range stops below f17.01.49, so it looks safe — our inference, not a vendor statement
abbott-assurity-mri (same device)

Remediation

No guide yet

No remediation guidance has been written for this CVE yet, and no fixed version or vendor advisory is recorded against it.

References (2)

VDB Entryx_refsource_BIDThird Party Advisory
http://www.securityfocus.com/bid/100523
x_refsource_MISCThird Party AdvisoryUS Government Resource
https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01

CVSS

Confidentiality
NONE
Integrity
NONE
Availability
COMPLETE
Vector?

Quick Info

StatusMODIFIED
CWE
CWE-920

Get alerted

Stop reading CVE feeds looking for your own stack

Track the products you run and VulnTracker tells you when a new CVE affects them — with what is known about fixing it, and nothing about the rest.

Free plan, no card