CVE-2015-1012

N/A
Published: March 25, 2019 at 06:20 PM UTC
Updated: June 17, 2026 at 12:21 AM UTC
3 products affected across 2 vendorsLifecare Pca Infusion System FirmwareLifecare Pca Infusion SystemLifeCare PCA Infusion System

Using LifeCare PCA Infusion System by Hospira?

Get instant alerts when new vulnerabilities affect LifeCare PCA Infusion System - before attackers find them.

Real-time alertsTrack your software stackFree plan available
Severity
N/A
Exploit prediction
EPSS 0.8%

Description

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.

Affected

Our own answer: ranges merged, affected separated from platform-only, with confidence and review flags.

v ≤ 5.0
pfizer-lifecare-pca-infusion-system (same device)

Remediation

No guide yet

No remediation guidance has been written for this CVE yet. What the record does show:

References (1)

x_refsource_MISCMitigationThird Party AdvisoryUS Government Resource
https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01

CVSS

Confidentiality
PARTIAL
Integrity
NONE
Availability
NONE
Vector?

Quick Info

StatusMODIFIED
CWE
CWE-312CWE-200