CVE-2014-5406

Hospira LifeCare PCA Infusion System

HIGH
7.6
CVSS Score
Published: July 6, 2015 at 07:10 PM UTC
Updated: June 17, 2026 at 12:11 AM UTC
4 products affected across 1 vendorLifecare Pcainfusion FirmwareLifecare PCA3Lifecare PCA5

Using N/A by N/A?

Get instant alerts when new vulnerabilities affect N/A - before attackers find them.

Real-time alertsTrack your software stackFree plan available
Severity
HIGH7.6
Exploit prediction
EPSS 1%

Description

The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.

Affected

Our own answer: ranges merged, affected separated from platform-only, with confidence and review flags.

v ≤ 5.0
hospira-lifecare-pca3 (environment) · hospira-lifecare-pca5 (environment)
v ≤ 5.0
Vendor states this is fixed in 7.0 (unaffected)
All versions
All versions

Remediation

No guide yet

No remediation guidance has been written for this CVE yet, and no fixed version or vendor advisory is recorded against it.

CVSS

7.6HIGHicscert
Vector?

Quick Info

StatusMODIFIED
CWE
CWE-345

Get alerted

Stop reading CVE feeds looking for your own stack

Track the products you run and VulnTracker tells you when a new CVE affects them — with what is known about fixing it, and nothing about the rest.

Free plan, no card